Alignment
Wash Head — Automatic Depth Alignment
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-62
September 2016
Wash Head — Automatic Depth Alignment
Follow the steps below to perform the automatic depth alignment.
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<4> Alignment
.
OR
Enter
4
in the
Option No.
field and press
[Enter]
.
3
Use the
<Page Up>
and
<Page Down>
buttons to view additional Utilities
options.
Select
<4> Cuvette Wash
from the Functional Areas dialog box.
OR
Enter
4
in the
Area No.
field and press
[Enter]
.
4
Select
<3> Automatic Depth Alignment
.
OR
Enter
3
in the
Procedure No.
field and press
[Enter]
.
5
Complete steps 1-3 as described on the screen, selecting
<Continue>
after
performing each step.
OR
Select
<Stop>
to exit. Select
<Close>
to return to the Alignment screen or
select
<Repeat>
to restart the alignment procedure.
6
Select
<Yes>
from the Alignment dialog box to save the alignment data to
a temporary file.
OR
Select
<No>
to return to the Alignment screen without saving the
alignment data.
7
Select an icon from the menu bar.
Select
<Yes>
from the alignment dialog box to save the new alignment data
to the database.
[F1] Prev Align
allows the system to overwrite this
alignment data with the previous alignment data.
OR
Select
<No>
to restore the previous alignment data.