Reference Interval Setup
Deselecting a Default
IMMAGE 800 Operations Manual A11403
System Setup
September 2016
Page 5-21
5
Deselecting a Default
When the instrument status is in
Standby
, proceed with the steps below to deselect the
default interval and range.
Clearing Intervals/Ranges
Intervals and ranges can be cleared from either the Reference Interval/Critical Ranges
dialog box or the Reference Interval/Critical Ranges screen.
When the instrument status is in
Standby
, proceed with the steps below to clear
intervals and ranges from the Reference Interval/Critical Ranges dialog box.
Step
Action
1
From the Setup screen, select
<4> Reference Interval
2
Select the button beside the desired chemistry/calculation. Multiple
chemistries/calculations may be chosen.
3
Select
<Define Edit>
4
Select a field on the interval/range line that is the current default.
5
Select
Select Default [F3]
.
The "
>
" symbol is removed from the far left of the line.
6
Select another sample type and/or sex to move to other definitions within
the same chemistry/calculation. Repeat Steps 4-5.
OR
Select
Prev Chem [F9]
or
Next Chem [F10]
for other chemistries/
calculations. Repeat Steps 4-5.
OR
Select any icon from the menu bar to exit the screen.
Step
Action
1
From the Setup screen, select
<4> Reference Interval
2
Select the button beside the chemistry/calculation to be cleared. Multiple
chemistries/calculations can be chosen.
3
Select
<Clear Ranges>
.
4
Select
<OK>
to clear all the intervals and ranges for the chemistry(ies)/
calculation(s) selected.
OR
Select
<Cancel>
to return to the Reference Interval/Critical Ranges dialog
box without clearing intervals and ranges.