Accessing Instrument Commands
The Stop Command and Cuvette Maintenance
System Status/Instrument Commands
IMMAGE 800 Operations Manual A11403
Page 11-8
September 2016
The Stop Command and Cuvette Maintenance
The following steps are recommendations for maintenance of the cuvettes after the
Stop
command is selected.
Table 11.4
Recommended Maintenance After Stop Command Selected
After Stop is selected
during a run and before
Run is selected...
the cuvettes...
immediately
(<1 hour)
do not require maintenance.
less than 24 hours later
should be washed (Refer to CHAPTER 10,
Utilities
,
As-Indicated Maintenance, "Washing Cuvettes".)
more than 24 hours later
should be replaced. (Refer to CHAPTER 10,
Utilities
,
As-Indicated Maintenance, "Replacing Cuvettes".)