Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-36
September 2016
1800
Fatal Results
Calculation Task
Error
WARNING: Will STOP
Run
Record what is displayed on pop-up
screen.
CALL BECKMAN COULTER
CLINICAL SUPPORT
1801
Calibration
Failed for
Reagent
WARNING:
Repeat calibration.
2001
Unrecognized
Bulk Name
WARNING: Will STOP
Run
Verify that buffer and diluent positions
are defined correctly.
2002
Size of memory
allocation
mismatch.
No Chemistry
Protocol
Available for this
dilution.
FATAL:
Check Sample Type for programmed
chemistries.
2003
Buffer Detected
but Not Defined
WARNING:
Define buffer positions for all loaded
buffers.
2004
Buffer Defined
but Not Detected
WARNING:
Load buffers for all defined buffer
positions.
2005
Diluent Detected
but Not Defined
WARNING:
Define diluent positions for all loaded
diluents.
2006
Diluent Defined
but Not Detected
WARNING:
Load diluents for all defined diluent
positions.
2007
Invalid Card
Type Received
from MPC
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
2008
Reagent Handler
Invalid
Parameters
WARNING: Will STOP
Run
CALL BECKMAN COULTER
CLINICAL SUPPORT
2201
Illegal Protocol
Tree Address
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
2202
Scheduler Fatal
Error
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
2204
Corrupt Cycle
Stack
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(15 of 21)