Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-38
September 2016
2222
Task Fatal Error
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
2223
Error with
Scheduler Queue
WARNING: Will STOP
Run
CALL BECKMAN COULTER
CLINICAL SUPPORT
2224
Illegal Queue
Message
Received
WARNING: Will STOP
Run
CALL BECKMAN COULTER
CLINICAL SUPPORT
2225
Error in Cycle
Count
Conversion
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
2402
Reaction
Compartment
Temperature Out
of Range Low
WARNING: Will STOP
Run
1. Verify cover is closed.
2. Check Status Monitor for stable
reaction temperature in acceptable
range.
2403
Reaction
Compartment
Temperature Out
of Range High
WARNING: Will STOP
Run
1. Verify cover is closed.
2. Check Status Monitor for stable
reaction temperature in acceptable
range.
2404
Reagent
Compartment
Temperature Out
of Range Low
WARNING:
1. Verify reagent fans are working.
2. Verify cover is closed.
3. Check Status Monitor for stable
reagent compartment temperature
in acceptable range.
2405
Reagent
Compartment
Temperature Out
of Range High
WARNING:
1. Verify reagent fans are working.
2. Verify cover is closed.
3. Check Status Monitor for stable
reagent compartment temperature
in acceptable range.
2406
Excessive
temperature
difference in the
reaction chamber
WARNING:
1. Verify reagent fans are working.
2. Verify cover is closed.
3. Check Status Monitor for stable
reagent compartment temperature
in acceptable range.
2408
Analog Ground
Out of Range
Low
WARNING:
Observe Status Monitor
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(17 of 21)