Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-72
September 2016
Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
The instrument status must be in
Standby
in order to proceed with the steps below to
replace the Sample Probe or the Reagent Probe.
Step
Action
1
Disable all Electro-mechanical motions. (Refer to Troubleshooting,
Callable Diagnostics, "Home All and Disable All" in this chapter.)
2
Remove the left access cover (1) of the Crane Assembly by removing the
two Phillips screws on the left side.
Loosen the Phillips screws on the right side (2) of the assembly. (Refer to
Figure 10.36.)
Figure 10.36
(1 of 6)
1
2
A014076P.EPS