Callable Diagnostics
Introduction
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-49
10
Callable Diagnostics
Introduction
The Callable Diagnostics option provides procedures directed toward identifying and
isolating specific IMMAGE 800 malfunctions. This section describes:
• Home All
• Disable All
• Sample Stir Motor
• Reagent Stir Motor
The instrument status must be in
Standby
in order to proceed with the procedures
described below.
Home All and Disable All
Follow the steps below to home the system or to disable all instrument motors.
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<3> Diagnostics
.
OR
Enter
3
in the
Option No.
field.
3
Select
<2> Callable Diagnostics
from the Diagnostics dialog box. (Refer
OR
Enter
2
in the
Option No.
field and press
[Enter]
.
4
Select
<1> Electro-Mechanical Motion
from the Callable Diagnostics
screen.
OR
Enter
1
in the
Option No.
field and press
[Enter]
.
5
Select…
to…
Call. Diag [F1]
return to the Callable Diagnostics
screen.
Home All [F3]
home the system.
Disable All [F4]
disable all instrument motors.