Antigen Excess (AGXS) Testing
Introduction
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-30
September 2016
Antigen Excess (AGXS) Testing
Introduction
Antigen excess (AGXS) testing can be enabled or disabled for an individual chemistry.
The enabling or disabling of AGXS from Sample Options applies only to the current
sample.
If AGXS testing is applicable to a chemistry, a check box appears beside that
chemistry in the Sample Options dialog box.
AGXS Testing Default
The default for AGXS testing is defined in Setup and indicated by the check box.
(Refer to Table 7.6.)
Enabling/Disabling AGXS Testing
Follow the steps below to enable or disable AGXS testing.
Additional Information
Refer to CHAPTER 3,
Theory of Operation
, Antigen Excess Testing for more general
information on antigen excess.
Refer to CHAPTER 5,
System Setup
, Configuring Antigen Excess Testing for more
information on defining the default for AGXS testing.
Table 7.6
AGXS Testing
When the check
box is...
AGXS testing is...
And on the Program Sample
screen...
checked,
enabled for the chemistry.
Every time the test is run
it will be tested for
antigen excess.
a small green box appears on the
chemistry button on the chemistry
menu.
unchecked,
disabled for the
chemistry. The chemistry
will not be tested for
antigen excess.
there is no green box on the chemistry
button on the chemistry menu.
Step
Action
1
Select
Sample Options [F3]
from the Program Sample screen.
2
Choose the AGXS check box beside the desired chemistry.
3
Select the check box to toggle it to "checked" or "unchecked".