Selecting Replicates
Introduction
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-29
7
Selecting Replicates
Introduction
A Sample Replicate number can be entered in the
Sample Replicate
field. The
Sample Replicate number defines the number of times all tests selected for the sample
will be repeated. (1-9 Sample Replicates are available.) It applies only to the current
sample.
The default number of sample replicates defined in Setup is displayed in the
System
Replicates
field. This field is for display only and cannot be accessed, except from
Setup.
A test replicate number can be entered for each test that is selected for a sample. The
test replicate number defines the number of times an individual test will be repeated.
(1-9 test replicates are available). It applies only to the current sample.
Entering Sample Replicates
Follow the steps below to enter a Sample Replicate number.
Entering Test Replicates
Follow the steps below to enter a test replicate number.
Additional Information
Refer to CHAPTER 5,
System Setup
, Default Setup, for more information on defining
System Replicates.
NOTICE
The test replicates function is available only when the Sample Replicate is set to 1.
Step
Action
1
Select chemistry(ies) from the Program Sample screen.
2
Select
Sample Options [F3]
.
3
Choose the
Sample Replicate
field.
4
Type the number of replicates for the sample (1-9).
Step
Action
1
Select chemistry(ies) from the Program Sample screen.
2
Select
Sample Options [F3]
.
3
Choose the
Reps
field beside the desired chemistry.
4
Type the number of replicates for the test (1-9).