As-Indicated Maintenance
Washing Cuvettes
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-19
10
Washing Cuvettes
Washing the cuvettes may become necessary when performing maintenance or when
troubleshooting. To access cuvette washing, perform the following steps.
Replacing Syringes
To replace the reagent or sample syringes, perform the following steps.
12
Type the last cuvette number (39) in the
Cuvette
field to stop the wash in
the
Ending Range
field.
13
Select
<OK>
to start washing the cuvettes.
Step
Action
, continued
(3 of 3)
Step
Action
1
Select
Utilities
from the menu bar.
2
Select
<7> Wash Cuvettes
from the Utilities dialog box.
OR
Enter
7
in the
Option No.
field and press
[Enter]
.
3
Type the first cuvette number (1) to begin the wash in the
Starting Range
field and press
[Enter]
.
4
Type the last cuvette number (39) to stop the wash in the
Ending Range
field.
5
Select
<OK>
to start washing the cuvettes.
NOTICE
If
<Stop>
is selected to discontinue the cuvette wash, Homing of the
system is necessary prior to using the Sample Carousel Advance button.
OR
Select
<Cancel>
to return to Utilities dialog box without starting the wash.
Step
Action
1
Select
Utils
from menu bar.
2
Select
<3> Diagnostics
.
3
Select
<2> Callable Diagnostics
.
4
Select
<1> Electro-mechanical motion
.
5
Select
Home All [F3]
.
6
Select
Devices [F2]
.
7
Select either
<11> Sample syringe
or
<12> Reagent syringe
.
(1 of 2)