Loading/Clearing Buffers and Diluents
Loading a New Lot or Changing a Position
Reagents/Calibration
IMMAGE 800 Operations Manual A11403
Page 6-10
September 2016
Figure 6.6
Buffer/Diluent Status Dialog Box
Loading a New Lot or Changing a Position
The instrument status must be in
Standby
in order to proceed with the steps below to
change the lot or position of a buffer or diluent.
E010234S.EPS
Step
Action
1
Select
Rgts/Cal
from the menu bar.
2
Select
Buffer/Diluent [F3]
3
Select the options button
<
▼
>
beside the desired position.
4
Select the desired buffer or diluent to place in the position. Press
<OK>
.
OR
Select
<Cancel>
to return to the Reagents Status/Calibration Status screen
without selecting a buffer or diluent.
5
Enter the buffer or diluent lot number.
6
Repeat steps 3-5 for each buffer and diluent to be loaded.
7
Select
<OK>
to accept any changes made.
OR
Select
<Cancel>
to return to the Reagent Status/Calibration Status screen
without entering any changes.
8
Follow "Replacing the Same Lot," in this section, to place bottles on the
carousel(s).
NOTICE
Recalibration of affected chemistries may be necessary when buffer or
diluent lots are changed.