Database Recovery
Introduction
IMMAGE 800 Operations Manual A11403
System Power On/Off
September 2016
Page 4-5
4
Database Recovery
Introduction
In the event of a fatal system error or power outage, the system will attempt to recover
files and databases which may have been affected.
Fatal Errors
When a fatal error note is displayed, call Beckman Coulter for assistance.
Information Retained
The following information will be retained in the database after power is restored.
• Chemistry protocols
• Reagent and calibrator bar coded parameters
• Current calibration status for each reagent
• Calibration requests that were not completed
• Completed results
• Sample programming for incomplete results and samples that have not been run
• Quality control files
• All system software setup configurations
• Event logs
• All instrument hardware settings
Information Changed
The following information and/or screens will be changed after power is restored.
Table 4.1
Information Changed After Power is Restored
Information/Screen
Change
Dilution Segments dialog box
All segments go to 0
wells available
.
Sample Carousel Status monitor
Screen is blank.
Reagent Status/Calibration Status screen Screen is blank. Calibration requests are
deleted.
Sample Load List
Samples which were
In Process
, become
Incomplete
.
Instrument Status Monitor
Continuously updates all parameters.
Simulated calibrations
Calibration status changes to
Failed
.
Printing
Printouts will stop. From Results Recall,
request printouts of patient and control
reports if needed. Calibration reports
cannot
be recalled for printing.