Overview
Operating Without Host Communications
Sample Programming
IMMAGE 800 Operations Manual A11403
Page 7-82
September 2016
Operating Without Host Communications
Follow the steps below for routine operation of the IMMAGE 800 Immunochemistry
System using unidirectional mode.
Step
Action
1
Check and clear as necessary:
• sample racks
Check, clear, and replace as necessary:
• dilution segments
Check:
• reagents, buffers, and diluents
• wash solution volume
(Refer to Preparing for Programming/Running in this chapter.)
2
Request calibrations, if necessary. (Refer to CHAPTER 6,
.)
3
If samples are...
then...
bar coded,
program sample ID, chemistries, and any
additional information needed for each
sample. Load samples in any rack and
position.
not
bar coded,
program rack and position, sample ID,
chemistries, and any additional information
needed for each sample. Load samples in the
programmed racks and positions.
(Refer to Programming a Sample in this chapter and CHAPTER 2,
System
Description
, Sample Container Information.)
4
Select
Main
from the menu bar.
AND
Select
Run
.
5
Select
<OK>
in the Check Dilution Segments dialog box to start the run.
OR
Select
<Cancel>
to exit without starting the run.