Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-26
September 2016
E50
Dil N/A
Insufficient Diluent
Load a new diluent bottle, and verify
that diluent position is defined
correctly.
E51
Sample N/A
Insufficient Sample
Verify sufficent sample.
E52
Cal Fail
Calibration Failure
Repeat calibration.
E53
Pause
Pause
CALL BECKMAN COULTER
CLINICAL SUPPORT
E54
RGT Handler
Reagent Handler Error
CALL BECKMAN COULTER
CLINICAL SUPPORT
E55
WASH N/A
Insufficient Wash Fluid
1. Verify wash line is not pinched.
2. From Utilities menu, select
<8>
Fill Internal Wash Bottle
.
3. Rerun samples.
E56
No Model
No Model
CALL BECKMAN COULTER
CLINICAL SUPPORT
E57
RXN Temp
Reaction Temperature
Out of Range
1. Verify cover is in place.
2. Check Status Monitor for stable
temperature.
E58
ORLO
Low Result
Refer to IMMAGE Immunochemistry
Systems,
Chemistry Information
Manual
for appropriate measuring
range.
E59
ORHI
High Result
Refer to IMMAGE Immunochemistry
Systems,
Chemistry Information
Manual
for appropriate measuring
range.
E60
RXN Error
No Reaction, AGXS
failed
CALL BECKMAN COULTER
CLINICAL SUPPORT
E61
ORLO
No Reaction (excluding
Drugs)
Refer to IMMAGE Immunochemistry
Systems,
Chemistry Information
Manual
for appropriate measuring
range.
E62
RXN Error
No Reaction (Drugs)
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(5 of 21)