Alignment
Straightness of the Sample or Reagent Probe (from side)
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-56
September 2016
Figure 10.26
3
Remove the left access cover (1) of the Crane Assembly by removing the
two Phillips screws on the left side.
Loosen the Phillips screws on the right side (2) of the assembly. (Refer to
Figure 10.27.)
Figure 10.27
4
Rotate the assembly towards the appropriate compartment and raise the
probe guard to expose the probes.
Step
Action
, continued
(3 of 6)
A014075P.EPS
1
2
A014076P.EPS