Ranging Mode:
Unit ranging setting used during the Line Leakage test. Can be set to
Auto or Manual.
Continuous
: When active, the Continuous feature keeps the DUT powered from one
Line Leakage test to another, from one Functional Run test to another, and in between
Line Leakage and Functional Run test steps.
Scanner (Option):
Sets the Scanner channels to H (for Probe-HI connections), L (for
Probe-LO connections), or O (for open). This parameter will only appear when the
option 04 Scanner is installed. See section
8. Options
for more information.
PLC CONTROL (Option):
Selects the AC1000 memory location (M1-M7) to be used
during a Line Leakage or Functional Run test. This parameter will only be displayed
when Option 05 Remote Memory Send is installed. See section
8. Options
for more
information.
Defaults:
Pressing this soft key will default the test parameters to their predetermined
default values (refer to section
4.4.2. Default Test Parameters
). Once the “Default”
soft key is pressed it is necessary to complete the edit either by pressing the ENTER
key to accept the parameter overwrite or the EXIT key to escape from the edit and
return to the original values.
4.4.2. Default Test Parameters
In each of the Test Parameter Setting screens there is a “Defaults” soft key. When
this soft key is pressed the test will load with a set of predetermined default
parameters. The following table is a listing of the default parameters for each of the
different types of tests that are available in the 620L
.
The default
parameters are:
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...