97
When the Leakage-HI trip limit is set to 0, S2 is placed in position 2 which
disables the earth leakage test and connects the neutral of the DUT input to the
ground of the DUT.
03 Hipot/Ground Bond Interconnect
This option gives the 620L the capability to interface with an Associated
Research Hipot or Ground Bond tester. Three (3) additional input ports (HV,
Return, and CASE) and one (1) additional output port (Current) are added to the
back panel of the instrument, which should be used to connect the safety tester
to the 620L. For more information on the HV, Return, CASE, and Current
connections, please refer to section
3.2.2. Rear Panel Controls
.
The operator must also connect the included communication cables to the
Remote
Signal I/O’s of both the 620L and the electrical safety tester. These
cables will provide the communication path for the instruments to function
together.
With this option installed, the 620L and the electrical safety tester will operate in
a master/slave configuration. The test will be initiated from the front panel TEST
button of the 620L. Once the test is complete, the 620L will send a signal to the
electrical safety tester to begin any subsequent tests.
Note
: The PLC Remote
parameter in the electrical safety tester’s System
Menu screen must be turned ON for this option to work.
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...