187
Base Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA,
frequency 15Hz - 1MHz
Resolution
0.1uA, 1uA
Base Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 5
504uA ~ 999.9uA, 1000uA ~
8400uA,
frequency 15Hz - 1MHz
Resolution
0.1uA, 1uA
Base Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 6
1200uA ~ 8399uA, 8.40mA ~
20.00mA,
frequency 15Hz - 1MHz
Resolution
1uA, 0.01mA
Base Accuracy
15Hz < f < 100KHz : ±5% of
reading (0.01mA -20.00mA )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display (RMS)
(AC Only)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0uA ~ 7.8uA,
frequency 15Hz - 1MHz
Resolution
0.1uA