GPIB Queries and Responses
The 620L GPIB bus will not send any data to the controller without being queried. A
GPIB read command must be sent after a command string to retrieve any data from a
query command (?).
Rules for Sending Commands to the Instrument
The following conventions are used to describe the commands syntax for the 620L:
Braces ({ }) enclose each parameter for a command string.
Triangle brackets (< >) indicate that you must substitute a value for the
enclosed parameter.
The Pipe ( | ) is used to separate different parameter options for a command.
The command and the parameter data must be separated with a space.
Each command string should be terminated by the ASCII control code, New
Line (NL), (OAh) or the end of line (EOL) message for GPIB.
All commands that end with a question mark (?) are query commands and
required an IEEE-488 read command to retrieve the data from the device's
output buffer.
7.4.2. Test Execution Commands
The following commands are used to control actual output voltage and current from
the instrument. Please observe all safety precautions.
COMMAND
DESCRIPTION
TEST
Execute a Test
RESET
Abort a test in Process or Reset Failures
TEST
Starts the test sequence at the selected step loaded into memory (RAM).
RESET
Stop or abort a test. Also used to reset a latched failure condition.
7.4.3. File Editing Commands
The following commands are used to create or modify Test Setup Files.
Commands should be separated from parameters by a space.
If multiple parameters are entered, they should be separated by commas.
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...