196
Accuracy
15Hz < f <1MHz: ±(10% of
r 2uA)
Range 6
8000uA ~ 8399uA, 8.40mA ~
30.00mA, frequency DC, 15Hz
–
100KHz
Resolution
1uA, 0.01mA
Accuracy
15Hz < f <100KHz: ±(10% of
r 2counts)
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display (Peak)
(AC Only)
Range 1
0.0uA ~ 32.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2uA)
100KHz < f < 1MHZ : ±(10% of
r 2uA)
Range 2
7.8uA ~ 130.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2uA)
100KHz < f < 1MHZ : ±(10% of
r 2uA)
Range 3
31.5uA ~ 525.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2uA)
100KHz < f < 1MHZ : ±(10% of
r 2uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA,
frequency 15Hz - 1MHz
Resolution
0.1uA, 1uA
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2uA)
100KHz < f < 1MHZ : ±(10% of
r 2uA)
Range 5
504uA ~ 999.9uA, 1000uA ~
8400uA,
frequency 15Hz - 1MHz
Resolution
0.1uA, 1uA