128
Serial Number field, and if the user re-scans another barcode, the Data Storage Card
will replace the data in the Product Number field.
The RUN FILE selection gives the user the ability to automatically load and execute a
test file based on what barcode is scanned from the Perform Tests screen.
Using the RUN FILE feature will enable the instrument’s output once the barcode is
scanned. Do not touch the DUT at any time when using this feature in order to avoid
potential shock or serious injury.
To completely enable this feature, the user must name the desired test file for a
particular product the exact alpha-
numeric code that is on the product’s barcode label.
For example, if Product A has barcode “123456789”, then the test file that the user
would like to run when testing Product A should be named “123456789”. Upon
scanning the barcode, the 620L will immediately execute the test associated with that
barcode. The test file name is limited to 10 characters. However, if the user names a
test file with the maximum 10 characters, this function will still initiate a test when a
product’s barcode begins with those first 10 characters even if the barcode has more
than 10 characters.
When using certain features
of barcoding, the 620L’s response to the
TD? and RD x? commands will differ slightly to an 620L with the
standard USB/RS-232 interface. For all types of tests (LLT, RUN) two fields are
added to the end of the standard response when the Barcode I/P setting is set to
SERIAL#, PRODUCT# or SER/PROD. The first field contains the Serial Number
information and the second field includes the Product Number information. Both fields
are included regardless of which of these three modes are selected. The Data
Storage Card will simply substitute a “0” for the field if it is not applicable to the setting.
For example, if a user had their Barcode I/P setting set to SERIAL#, and scanned a
Serial Number with the value “123456789”, the TD? response for an LLT test could
be:
WARNING
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...