136
The above message appears when the USB Flash memory drive is write-protected.
There are two ways to configure the memory drive as write-protected. One is to set
the switch on the side of the memory dri
ve up (towards the drive’s connector). The
other is to configure it through the drive’s properties in Windows.
08 Leakage Current Offset Circuit
The 620L Line Leakage Tester allows for an offset of leakage current measurements
taken during a line leakage test. The purpose of the LLT Offset is to allow the unit to
zero out leakage current associated with the 620L instrument, the instrument leads
and any other integrated Associated Research, Inc. instruments. Since there is a
capacitive value associated with the Associated Research Inc. instruments, leakage
current due to this capacitance factors into the total leakage value displayed during a
line leakage test.
The 620L applies a formula for taking an offset into account:
Dis
played Current = √((Total Leakage Current)^2 – (Offset Current)^2)
The “Total Leakage Current” is the measure of the total leakage current on the system
including the 620L, other integrated Associated Research, Inc. instruments and the
device under test.
The “Offset Current” is the offset current value either specified by
the user or calculated automatically by the 620L unit. These two values are used to
calculate the actual displayed leakage current measurement. The 620L then
calculates the vector sum of the offset value and the total value so that the inherit
leakage of the Associated Research, Inc. instruments is excluded from the
measurements.
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...