203
reading (10.0mV - 999.9mV)
Range 2
7.8mV ~ 130.0mV, frequency
DC, 15Hz - 1MHz
Resolution
0.1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0mV - 999.9mV)
Range 3
31.5mV ~ 525.0mV, frequency
DC, 15Hz - 1MHz
Resolution
0.1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0mV - 999.9mV)
Range 4
126.0mV ~ 999.9mV,1000mV ~
2100mV, frequency DC, 15Hz -
1MHz
Resolution
0.1mV, 1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0mV - 999.9mV)
Range 5
504.0mV ~ 999.9mV, 1000mV
~ 8400mV, frequency DC, 15Hz
- 1MHz
Resolution
0.1mV, 1mV
Basic Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0mV - 999.9mV)
Range 6
1200mV ~ 8399mV, 8.40V ~
20.00V, frequency DC, 15Hz
–
100kHz
Resolution
1mV, 0.01V
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...