78
COMMAND
DESCRIPTION
VALUE
FL <
file number
>
File Load
file number
= 1-50
FD
File Delete
Active selected File
FD <
file number
>
File Delete by
Number
file number
= 1-50
FS
File Save
Active selected File
FSA {<
file
number
>,<
file
name
>}
File Save-as
file number
= 1-50,
file name =
Valid ASCII (1)
maximum 11 characters
FM {1|0}
File Memory Link
1=On, 0=Off
FM?
FN {<
file
number
>,<
file
name
>}
File New
file number
= 1-50,
file name =
Valid ASCII (1)
maximum 11 characters
SS <
step number
>
Step Select
step number
= 1-30
ADD
<test,
p1,p2,p3…>
Adds all parameters
SAR
Step Add RUN test
SAL
Step Add LLT test
SD
Step Delete
Active selected Step
SD <
step number
>
Step Delete by
Number
step number
= 1-30
SP <
prompt
message
>
Step Prompt Create
prompt message
= Valid ASCII
(1)
maximum 32 characters
SP
Step Prompt Delete
SF {1|0}
Step Fail Stop
1=On, 0=Off
(1) "Valid ASCII" is the character set that is available from the front panel LCD user
interface. Consisting of upper case alphabet (A-Z), numbers (0-9) and decimal point
(.), asterisk (*), dash (-), under bar (_), tilde (~) and space (SP).
FL <memory number>
Load a file by memory number from non-volatile memory into random access memory
RAM.
FD
Deletes the file loaded into RAM from the non-volatile memory space as well as RAM.
A new file must be loaded before and tests can be run.
The “FD” command writes to the non-volatile memory of the
instrument. The non-volatile memory has a limited write cycle
life. Therefore, for programmers who wish to send all
parameters before executing each test, the “FD” command should not be used.
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...