184
Resolution
1uA, 0.01mA
Base Accuracy
DC, 15Hz < f < 100KHz : ±5%
of reading (0.01mA -20.00mA )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display (RMS)
(AC+DC)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0uA ~ 7.8uA, frequency DC -
1MHz
Resolution
0.1uA
Additional Error
DC, 15Hz < f < 100KHz: add
±(2% of r 0.2% of
range) to base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 3
0.0uA ~ 31.5uA, frequency DC -
1MHz
Resolution
0.1uA
Additional Error
DC, 15Hz < f < 100KHz: add
±(2% of r 0.2% of
range) to base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 4
0.0uA ~ 126.0uA, frequency DC
- 1MHz
Resolution
0.1uA
Additional Error
DC, 15Hz < f < 100KHz: add
±(2% of r 0.2% of
range) to base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 5
0.0uA ~ 504uA, frequency DC -
1MHz
Resolution
0.1uA