201
Range 4
0.0mV ~ 126.0mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 5
0.0mV ~ 504mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 6
0.0mV ~ 999.9mV, 1000mV ~
1200mV, frequency DC
–
100KHz
Resolution
0.1mV, 1mV
Additional Error
DC, 15Hz < f < 100KHz : add
±(2% of r 0.2% of
range) to base accuracy
DISPLAY AUTO RANGE
Touch Voltage Display (RMS)
(AC Only)
Range 1
0.0mV ~ 32.0mV, frequency
DC, 15Hz
– 1MHz
Resolution
0.1mV
Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0mV - 999.9mV)
Range 2
28.0mV ~ 130.0mV, frequency
DC, 15Hz - 1MHz
Resolution
0.1mV
Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f <100KHz: ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...