186
Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 5
1800uA ~ 8400uA,
frequency 15Hz - 1MHz
Resolution
1uA
Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 6
8.00mA ~ 20.00mA,
frequency 15Hz
– 100kHz
Resolution
0.01mA
Accuracy
15Hz < f < 100KHz : ±5% of
reading (0.01mA -20.00mA )
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display (RMS)
(AC Only)
Range 1
0.0uA ~ 32.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 2
7.8uA ~ 130.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Base Accuracy
15Hz < f <30Hz: ±(3% of
r 5counts)
30Hz < f < 100KHz : ±(2% of
r 3counts)
100KHz < f < 1MHZ : ±5% of
reading (10.0uA - 999.9uA)
Range 3
31.5uA ~ 525.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...