121
If the Barcode I/P is set to PRODUCT#, scan the barcode once to input it into the
620L. The 620L will then search for a test file name that matches the product number
barcode string. If the 620L finds a match, it will load the file into RAM. When the
same product
number barcode is scanned a second time, the test will be executed
automatically. If 620L does not find a file name that matches the barcode string, the
unit will beep
– notifying the user that it did not find a matching file name. The test file
name is limited to 10 characters. However, if the user names a test file with the
maximum 10 characters, this function will still load a test file if the first 10 characters of
the product number match the file name.
If the Barcode I/P is set to SER/PROD, scan the serial number once to input it into the
620L. Next, scan the product number. From this point, 620L will
operate the same as when the Barcode I/P setting is set to
PRODUCT#.
The Autostart feature will not work with the SERIAL# setting.
The Autostart
feature will enable the instrument’s output once the product number
barcode is scanned a second time when in the PRODUCT# and SER/PROD modes.
Do not touch the DUT at any time when using this feature in order to avoid potential
shock or serious injury.
Ethernet Card Settings Commands and Companion Queries
Command
Name
Value
SIM {1|0}
SIM?
Set IP Mode
1=Manual, 0=Auto
(DHCP/BOOTP)
SIA <value>
SIA?
Set IP Address
Dotted decimal form. Ex.
192.168.1.50
SGA <value>
SGA?
Set Gateway IP
Address
Dotted decimal form
SSM <value>
SSM?
Set Subnet Mask
Dotted decimal form
SDN <value>
SDN?
Set Device Name
8 character max, must start with
a letter
MAC?
MAC Address Query
Example response:
00:20:4A:8B:B4:30
SBI {0|1|2|3|4}
SBI?
Set Barcode Input
0=Off, 1=Ser/Prod, 2=Serial#,
3=Product#, 4=Run File
SAS {1|0}
SAS?
Set Autostart
0=Off, 1=On
Communication Considerations
All of the above commands (excluding the query commands) will respond with
the 06 hex (6 decimal) Acknowledge (ACK) ASCII control code if the transfer
was recognized by the instrument.
WARNING
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...