190
Accuracy
DC : ±5% of reading (0.01mA -
20.00mA )
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Current Display
(DC Only)
Range 1
0.0uA ~ 32.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of r 3counts)
(>10uA)
Range 2
7.8uA ~ 130.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of r 3counts)
(>10uA)
Range 3
31.5uA ~ 525.0uA, DC
Resolution
0.1uA
Base Accuracy
DC : ±(2% of r 3counts)
(>10uA)
Range 4
126.0uA ~ 999.9uA, 1000uA ~
2100uA, DC
Resolution
0.1uA, 1uA
Base Accuracy
DC : ±(2% of r 3counts)
(>10uA)
Range 5
504uA ~ 999.9uA, 1000uA ~
8400uA, DC
Resolution
0.1uA, 1uA
Base Accuracy
DC : ±(2% of r 3counts)
(>10uA)
Range 6
1200uA ~ 8399uA, 8.40mA ~
20.00mA, DC
Resolution
1uA, 0.01mA
Base Accuracy
DC : ±5% of reading (0.01mA -
20.00mA )
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display
(DC Only)
Range 1
N/A
Resolution
N/A