11
Note
: The Adapter Box (universal US polarity) and line cord listed are
American. Other combinations of line cord and Adapter Box are available upon
request.
2.2. Installation
2.2.1. Work Area
Locate a suitable testing area and be sure you have read all
safety instructions for the operation of the instrument and
suggestions on the test area setup in section
1.4. Safety
.
Make sure the work area you choose has a three-prong grounded outlet capable of
supplying the necessary input current to the power source. Be sure the outlet has
been tested for proper wiring before connecting the instrument to it.
The mains plug is used as the disconnecting device and shall
remain readily operable. The socket-outlet shall be installed.
Near the equipment and shall be easily accessible.
2.2.2. Power Requirements
This instrument requires a power source of either 115 volts AC ± 10%, 50/60 Hz
single phase or 230 volts AC ±10%, 50/60 Hz single phase. Before applying power
verify that the instrument is set to the correct line voltage (115 or 230 volts). Adjust
the voltage select switch to 115 for use with a 115 volt input. Adjust the voltage select
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...