The 620L can also automatically take an offset reading. In order to do this, it is
recommended the user attach any components that should be offset from the total
measurement. This includes other Associated Research, Inc. instruments working as
a system with the 620L (for example, a MedTEST system or an SC6540 scanning
matrix). It is recommended the entire system be setup first before taking an automatic
reading to account for the leakage associated with all testing units. Do NOT connect
the DUT if implementing an offset. For details on how to connect another Associated
Research Inc. instrument to the 620L, please refer to Option 3 in this section. For
details on how to connect an SC6540 scanner to the 620L, please refer to Option 4 in
this section.
From the Line Leakage Parameter screen, use the arrow keys to scroll down and
highlight
the “Offset” parameter. A message will come up below the Offset parameter
box
: “Press TEST key for automatic offset setting.” Pushing the TEST key will allow
the 620L to automatically read the leakage of the system.