7.4.4. Test Parameter Editing Commands and Companion Queries
These commands are used to modify the test parameter within each step. These
commands require a parameter value to be included with the command. The
companion query command will read the parameter. The writing of the parameter
requires that the unit not be included with the value, only the numeric value should be
included with the command. Also, when the query commands are used the response
will not include the unit’s characters. Many of the commands will function the same
way for multiple test types; however, the input range may be different and therefore
used a different possible set of values.
COMMAND
NAME
TEST
TYPES
VALUE
EACDC {0|1|2}
EACDC?
Edit AC/DC Mode
LLT
0=ACDC, 1=AC, 2=DC
ECH <
value
>
ECH?
Edit Current-HI
RUN
0.00 - 15.00A
ECL <
value
>
ECL?
Edit Current-Lo
RUN
0.00 - 15.00A
ECTN {1|0}
ECTN?
Edit Continuous
Power
LLT
RUN
1=On, 0=Off
EDE <
value
>
EDE?
Edit Delay
RUN
LLT
0.0 - 999.9s
EDW <
value
>
EDW?
Edit Dwell
RUN
0.0 - 999.9s
EEM {1|0}
EEM?
Edit Extended
Meter
LLT
1=On, 0=Off
EG {1|0}
EG?
Edit Ground
LLT
1=Open, 0=Close
ELH <
value
>
ELH?
Edit Leakage-HI
RUN
0.000 - 10.00mA
ELL <
value
>
ELL?
Edit Leakage-LO
RUN
0.000 - 10.00mA
ELM {1|0}
ELM?
Edit Leakage
Mode
LLT
1=Peak, 0=RMS
EM
{0|1|2|3|4|6|8|9}
EM?
Edit Meas-Device
LLT
0=UL544NP
1=UL544P
2=IEC60601, UL2601
3=UL1563
4=IEC60990 FIG4-U2,
IEC 60950
6=IEC60990 FIG5-U3
8=External
9=Frequency Check
EN {1|0}
EN?
Edit Neutral
LLT
1=Open, 0=Close
EPFH <
value
>
EPFH?
Edit PF-HI
RUN
0.000 - 1.000
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...