DISPLAY (CONT.)
Resolution
0.01mA
Accuracy
DC, 15Hz < f < 100KHz : ±5%
of reading (0.01mA -20.00mA )
Touch Current Display (Peak)
Range 1
0.0uA ~ 32.0uA, frequency DC -
1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of r 2uA)
15Hz < f < 1MHZ : ±10% of
r 2uA
Range 2
28.0uA ~ 130.0uA, frequency
DC - 1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of r 2uA)
15Hz < f < 1MHZ : ±10% of
r 2uA
Range 3
120.0uA ~ 550.0uA, frequency
DC - 1MHz
Resolution
0.1uA
Accuracy
DC : ±(2% of r 2uA)
15Hz < f < 1MHZ : ±10% of
r 2uA
Range 4
400uA ~ 2100uA, frequency DC
- 1MHz
Resolution
1uA
Accuracy
DC : ±(2% of r 2uA)
15Hz < f < 1MHZ : ±10% of
r 2uA
Range 5
1800A ~ 8500uA, frequency DC
- 1MHz
Resolution
1uA
Accuracy
DC : ±(2% of r 2uA)
15Hz < f < 1MHZ : ±10% of
r 2uA
Range 6
8.0mA ~30.0mA, frequency DC
– 100KHz
Resolution
0.01mA
Accuracy
DC : ±(2% of r 3counts)
15Hz < f < 100KHz : ±10% of
r 2counts