205
Additional Error
15Hz < f < 30Hz : add ±(2% of
r 0.2% of range) to
base accuracy
30Hz < f < 100KHz: add ±(2% of
r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 5
0.0mV ~ 504mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Additional Error
15Hz < f < 30Hz : add ±(2% of
r 0.2% of range) to
base accuracy
30Hz < f < 100KHz: add ±(2% of
r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 6
0.0mV ~ 999.9mV, 1000mV ~
1200mV, frequency DC
–
100KHz
Resolution
0.1mV, 1mV
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
DISPLAY AUTO RANGE
Touch Voltage Display
(DC Only)
Range 1
0.0mV ~ 32.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 2
28.0mV ~ 130.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 3
120.0mV ~ 525.0mV, DC
Resolution
0.1mV
Accuracy
DC : ±(2% of r 3counts)
(>10.0mV)
Range 4
400mV ~ 2100mV, DC
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...