Fatal Error
If the instrument has a recognizable internal component failure and the TEST button is
pressed, the Fatal Error screen will appear as follows:
This type of failure permanently locks the instrument in the “Fatal Error” mode and
requires that the instrument be serviced by Associated Research, Inc. The following
Fatal Error identification numbers will represent the type of failure that has occurred:
Error Code 9001
will appear on the display if the instrument has a recognizable
internal component failure.
Error Code 9002
will appear on the display if the instrument’s System Data or the
Model/Option/Serial Number Data are corrupted and do not match the settings.
Error Code 9003
will appear on the display if the instrument’s Calibration Data is
corrupted.