9.2.7. Calibration of LLT MD V-Offset
Upper Right Screen
Load/Meter
Lower Right Screen
Prompt
Lower Right Screen
Prompt (Cal. Active)
Disconnect power to the
DUT I/P input.
Disconnect Probe-HI+
and Probe-LO-.
When the leads are
disconnected, press TEST
to start the calibration
process.
Automatic Calibration
Processing.
Press the TEST button to activate the calibration. After the calibration is activated, the
lower right screen will change to the text indicated in the table above (
Cal. Active
).
This calibration is automatic and does not require data entry.