197
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2uA)
100KHz < f < 1MHZ : ±(10% of
r 2uA)
Range 6
1800uA ~ 8399uA, 8.40mA ~
30.00mA,
frequency 15Hz - 1MHz
Resolution
1uA, 0.01mA
Base Accuracy
15Hz < f <100KHz: ±(10% of
r 2counts)
DISPLAY FIXED RANGE (<6% OF RANGE)
Touch Current Display (Peak)
(AC Only)
Range 1
N/A
Resolution
N/A
Additional Error
N/A
Range 2
0.0uA ~ 7.8uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 3
0.0uA ~ 31.5uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Range 4
0.0uA ~ 126.0uA,
frequency 15Hz - 1MHz
Resolution
0.1uA
Additional Error
15Hz < f < 100KHz : add ±(2%
of r 0.2% of range) to
base accuracy
100KHz < f < 1MHZ : add ±(2%
of r 0.5% of range) to
base accuracy
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...