212
Range 5
1800mV ~ 8400mV, frequency
DC, 15Hz - 1MHz
Resolution
1mV
Accuracy
15Hz < f <1MHz: ±(10% of
r 2mV)
Range 6
8000mV ~ 8399mV, 8.40V ~
30.00V, frequency DC, 15Hz
–
100KHz
Resolution
1mV, 0.01V
Accuracy
15Hz < f <100KHz: ±(10% of
r 2counts)
DISPLAY FIXED RANGE (>6% OF RANGE)
Touch Voltage Display (Peak)
(AC Only)
Range 1
0.0mV ~ 32.0mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2mV)
100KHz < f < 1MHZ : ±(10% of
r 2mV)
Range 2
7.8mV ~ 130.0mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2mV)
100KHz < f < 1MHZ : ±(10% of
r 2mV)
Range 3
31.5mV ~ 525.0mV,
frequency 15Hz - 1MHz
Resolution
0.1mV
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2mV)
100KHz < f < 1MHZ : ±(10% of
r 2mV)
Range 4
126.0mV ~ 999.9mV, 1000mV ~
2100mV,
frequency 15Hz - 1MHz
Resolution
0.1mV, 1mV
Base Accuracy
15Hz < f < 100KHz : ±(10% of
r 2mV)
100KHz < f < 1MHZ : ±(10% of
r 2mV)