107
The following Test Parameter Editing GPIB commands may be used to control
the memory send selection.
COMMAND
NAME
TEST
TYPES
VALUE
EPM n, n=1-7
EPM?
Edit PLC Memory
Send
LLT
RUN
1=M1, 2= M2, 3=M3,
4=M4, 5=M5, 6=M6,
7=M7
The ADD command can also be used. The memory selection should be inserted
as the last parameter and should use full text as indicated on the LCD screen
such as M1 or M2.
EPM
Sets the AC1000 memory location to be used during a Line Leakage or
Functional Run test.
06
Ethernet Card
The Ethernet Card option provides RS-232 and Ethernet communication
interfaces, as well as barcode scanning capability.
The Ethernet Card has three input/output ports, shown in the following figure:
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...