63
5.3. Performing a Test
1. As instructed in section
4.2. System Setup
, select a file and step that is
suitable for the test you would like to perform.
2. Attach the appropriate DUT to the instrument (refer to section
5.1. Instrument
Connections
for more information).
3. Press the TEST button.
4. The instrument will now perform the test or connected sequence of tests. If the
test is started from any step other than 01 when you push RESET or TEST it
will always return to the originally selected step.
5. If a Prompt is embedded within a step, the test will pause at that step and
display the Prompt as a pop-up message. An example of a Prompt will appear
as follows:
6. In order to clear the Prompt and continue the test, push the TEST button. The
highlighted area of the screen remains on the test that is being performed until
it is finished. On the right half of the screen the appropriate metering will
display for the type of test that is being performed (refer to section
5.4. Perform
Tests Metering
for more information). When the step is complete the
highlighted area will sequence to the next step and the new metering display
will appear (unless there is a prompt embedded within the step). When the
highlighted area moves to the next step the previous step area will display the
results of the test just performed.
7. At the end of the test sequence the right side of the screen will display a
flashing PASS if all the tests have completed successfully. The screen will
display a flashing FAIL if any of the tests did not complete successfully. If a
test fails the right side of the screen will indicate the step that failed and which
parameter of the step failed.
Содержание 620L
Страница 99: ...93 ...
Страница 105: ...99 The following figure shows how to connect the 620L to an Associated Research HypotULTRA III tester ...
Страница 106: ...100 The following figure shows how to connect the 620L to an Associated Research Omnia 8104 tester ...
Страница 123: ...117 From this screen four parameters may be accessed MAC Address Barcode I P Autostart and previous page ...
Страница 125: ...119 After the barcodes are scanned press TEST to initiate the test sequence Pressing RESET will abort the TEST sequence ...