13 Measuring
C76IB003EN-E
79
The progress bar shows the progress of the measurement with a growing
red bar and the message "Measuring".
During measurements the last 1, 2 or 3 digits might be gray colored. This
means that the set temperature has not been reached yet. In that case, only
the black colored digits are valid. The number of valid (black colored) digits
increases during measurement progress until all digits are black (when using
equilibrium mode).
When the measurement is finished, the progress bar turns green, the
message "Finished" is displayed and an acoustic signal is given. The
"Density Condition" output field shows the message "valid".
The result values are frozen. The display can be unfrozen by tapping the
button.
The result values are saved in the data memory and can be viewed, printed,
exported or deleted.
5. Measure the next sample or clean and dry the measuring cell.
Related topics
Safety Instructions: Chapter 2
Mounting the Hoses: Chapter 7.4
Measuring Methods: Chapter 12.1
Using the "Sample List" Mode: Chapter 13.3
Filling Samples: Chapter 13.4
Cleaning and Drying the Measuring Cell: Chapter 14.1
Handling the Measurement Data: Chapter 15
Measuring at High Humidity/Low Temperature Conditions: Appendix C.1
Measuring at Low/High Temperatures: Appendix C.2
Degassing Samples: Appendix B.1