12 Defining and Using Methods
C76IB003EN-E
61
9. Document the calibration procedure in a calibration protocol which contains
the operator’s name, date, place, description of the calibration procedure,
results and the calibration certificate of the liquid density standard.
Related topics
Performing Density Checks: Chapter 11.2.2
Performing an Air/Water Adjustment: Chapter 11.3.1
Cleaning and Drying the Measuring Cell: Chapter 14.1
Defining and Using Methods: Chapter 12
Performing Measurements: Chapter 13.5
12 Defining and Using Methods
12.1 Measuring Methods
Each method contains the following kind of information:
•
Instrument settings
•
Xsample settings and measuring module settings (if any module is
installed)
•
Layout of measuring data on the main screen
•
Measuring units
•
Parameter list for printout and data export
You can use the factory preset methods as they are or change them to suit your
needs (see Chapter 12.2). You can also create new methods (see Chapter 12.3).
Factory preset methods
The DMA M is delivered with a set of 10 predefined methods covering the most
common applications. The measuring temperature for these 10 methods is set to
20 °C.
Density
•
Density, Specific Gravity, Density
Temperature, Density Condition, U-View
TM
•
General purpose method