18 Communication with External PC and LIMS
112
C76IB003EN-E
start
or
start method_number
or
start method_number
magazine_position
wrong magazine position
An Xsample sample filling module is
active and the magazine position
parameter is not valid (e.g. smaller
than 1 or larger than the magazine
size).
sample storage full
The data memory and sample list
are full. Clear the data memory and
sample list or activate the automatic
clearing option (see Chapter 15.1).
heating
The Heating Attachment is heating,
so no measurement can be started.
abort
measurement aborted
The measurement was aborted.
measurement not started
No measurement was started.
already aborting
The measurement is already being
aborted.
getdatahead
Density;Specific Gravity;Density
Temperature;Density Condition
Example response with the default
settings.
no data available
No measurement has yet been
finished and so no data is available.
getdataunit
g/cm³;-;°C;-
Example response with the default
settings.
no data available
No measurement has yet been
finished and so no data is available.
getdata
0.000892;0.000894;20.001;valid Example response with the default
settings.
no new data available
No new measurement has yet been
finished and so no new data is
available.
finished
Measurement not started
No measurement has yet been
started.
Measurement not finished
The measurement is in progress
now.
Measurement finished
The measurement was finished.
getrawdata
1.102689;19.999;20.000:8
Example response.
getmethodname
method name: Density, 0
Example response.
getid
serial number: 80000000
DMA 4500M V1.70.6534.40
protocol version: 2.00
Example response.
set temperature xx.xxx
accepted
The command was accepted and
the density temperature was set to
the given value.
wrong parameter value
The given value was not accepted
(e.g. value is out of specifications).
Command
Response
Description