Appendix A: Technical Data
114
C76IB003EN-E
Appendix A: Technical Data
A.1
Measuring Performance
Table A.1: Technical data of measuring performance
DMA 4100 M
DMA 4500 M
DMA 5000 M
Measuring range
0 to 3 g/cm
3
Repeatability density (s. d.)
0.00005 g/cm
3
0.00001 g/cm
3
0.000001 g/cm
3
Repeatability temperature (s. d.)
a
0.02 °C
0.01 °C
0.001 °C
Measuring time
b
approx. 30 sec.
approx. 30 sec.
approx. 40 sec.
Measuring temperature
c
0 °C to 90 °C (32 to 194 °F)
from DCB version 1.000.009: 0 °C to 95 °C (32 to 203 °F)
Pressure range
d
0 to 10 bar (0 to 145 psi)
Sample volume
approx. 1 mL
Ambient air pressure sensor
yes
Full range viscosity correction
yes
Reference oscillator
yes
Automatic bubble detection
yes
Visual check of the measuring cell
Inspection
window
Camera
Camera
a.
While the temperature repeatability is constant for the whole temperature range (from 0 °C to 95 °C), the temperature
accuracy is not constant for the whole range. For temperatures above 90 °C, the temperature accuracy is typically
between 0.01 °C and 0.05 °C.
b.
After full temperature equilibrium
c.
Cooling down further than 20 °C (68 °F) below ambient temperature only with external cooling using the cooling kit
d.
For the measuring cell together with the injection adapters DMA/CarboQC (Mat. No. 19680) in the temperature range
of 0 to 50 °C (32 to 122 °F). For temperatures higher than 50 °C (122 °F), the maximum allowed pressure is 5 bar
(72.5 psi). See also Appendix C.2.
Table A.2: Typical accuracy
a
of concentration measurements (aqueous solutions, % w/w)
DMA 4100 M
DMA 4500 M
DMA 5000 M
Ethanol
0.05
0.025
< 0.01
Extract/Sugar
0.025
0.015
< 0.01
HCl
0.02
0.01
< 0.01
NaOH
0.04
0.02
< 0.01
H
3
PO
4
0.06
0.03
< 0.01
HNO
3
0.07
0.035
< 0.01
H
2
SO
4
0.05
0.025
< 0.01
a.
The accuracy is not the same over the whole concentration range. The values given are typical values for orientation
purposes.