10 Defining General Settings
42
C76IB003EN-E
10.1.3
Input Units
In this menu, you can define the unit for every temperature value which you enter
during operation.
1. Tap <Menu> and select
"Setup > Control Panel > Input Units"
.
2. Select the input unit for temperatures and tap <OK>.
10.1.4
Air Pump Settings
You can set the timeout after which the air pump is stopped and the condition on
which the air pump is stopped.
1. Tap <Menu> and select
"Setup > Measuring System Settings > Air
Pump"
.
2. Use the check box to activate/deactivate the option:
•
Air pump stops automatically when density value is stable.
•
Air pump stops automatically after ... s.
In case both options are activated and the air density value gets not stable
within the period of time that is defined with the second option, the air pump
stops nevertheless after this period of time.
3. Tap <OK> to save the changes.
10.1.5
Saving a Camera Picture (DMA 4500/5000 M Only)
You can define when the instrument automatically saves a picture of the
measuring cell.
1. Tap <Menu> and select
"Setup > Measuring System Settings > Camera"
.
2. Use the radio buttons to select one of the options:
Automatically save a picture ...
•
of each measurement
•
only if a measuring error is detected
•
never
3. Tap <OK> to save the changes.
10.1.6
Network
If you connect the instrument to your local network, you can use a network printer
for printouts and request the instrument system information and instruction
manual via any PC in the network.
1. Tap <Menu> and select
"Setup > Control Panel > Network"
.