12 Defining and Using Methods
68
C76IB003EN-E
To define formula parameters
1. Tap <Menu> and select
"Methods > Method Settings > "Method Name" >
Formula Parameters"
.
2. Select an API Product Group and API Input Quantity.
3. Select the special adjustment for the Canadian Excise Alcohol Table.
4. Tap <OK>.
12.2.7
Defining Quick Setting Parameters
Quick settings function
The quick settings function allows you to easily access different parameters
concerning your measurement without changing the current method.
The quick settings comprise several measuring settings or functions and are
available in all sample list modes. See Appendix G for an overview on the
parameters that can be defined as quick settings.
Depending on the sample list mode, different quick setting parameters are set by
default:
•
If the "No Sample List" mode is used: Sample name, measurement type
("Type") and "Density Temperature"
•
If the "Sample List" mode is used: Measurement type and "Density
Temperature"
Additional parameters can be added to your quick settings for every method.
User-defined data fields (see Chapter 10.3) and user functions of the "Constant"
type (see Chapter 16.2) will be automatically added to your quick settings.
If you have connected one or more Anton Paar measuring modules to your
master instrument, the range of parameters that can be added to your quick
settings will be extended. See the instruction manual of your measuring module
for information on the available parameters.
To set a quick setting parameter
1. Tap <Menu> and select
"Methods > Method Settings > "Method name" >
Quick Settings Management"
.
2. Tap the insert field in the free line.
3. Use the "Quantity" drop-down box to select the desired parameter and tap
<OK>.
4. Tap <OK> to save the settings.