7 Installing the Instrument
26
C76IB003EN-E
If the connections are leaking, the plunger will not move. In this case, repeat
the mounting of the adapters.
7.4
Mounting the Hoses
To connect the waste vessel
1. Cut a piece of approx. 250 mm (10 in) length from the silicone hose
contained in the accessory box.
2. Attach an adapter Luer cone (from the accessory box) to one end of the
silicone hose and insert it into the rear injection adapter of the DMA M.
3. Insert the other end of the silicone hose into the hole of the closed waste
vessel cap.
Fig. 7 - 2
Connecting the waste hose
To connect the silicone hose at the air pump outlet
1. Cut a piece of approx. 250 mm (10 in) length from the silicone hose
contained in the accessory box.
2. Attach the silicone hose to the air pump outlet.
3. Attach an adapter Luer cone (from the accessory box) to the other end of the
silicone hose.
WARNING
Liquids leaking from the instrument may cause injuries and risk of fire.
•
Only use the supplied hose and waste vessel if their materials are
resistant to the samples and cleaning liquids that you are going to inject.
•
If the supplied materials are not resistant, use other parts made of
appropriate material.