6 Functional Components
22
C76IB003EN-E
Quick access buttons
To open the diagnosis list.
The general instrument status as well as all measuring errors that
have occurred during the measurements of the currently active
sample list are described in this list. The button changes its
appearance depending on the current error status:
With green check:
The general instrument status and the error status of all measured
samples of the current sample list are OK.
With yellow warning sign:
•
The instrument (or system) has a minor problem (e.g. an air or
water check is overdue, there is a printer problem etc.).
•
One or more samples of the currently active sample list had a
filling error.
With red lightning sign:
•
The instrument (or system) has a major problem that needs to
be fixed before you continue with measurements (e.g. sample
changer is blocked).
•
One or more samples of the current sample list could not be
measured (e.g. measuring cell is partly empty so that it cannot
oscillate).
To reset the diagnosis button to the green check, confirm all error
messages by tapping on the <X> button on the right side of the
message.
The diagnosis button will also be reset to the green check symbol if
you delete the currently active sample list (see Chapter 13.3).
U-View
TM
: To open the live camera view of the measuring cell (DMA
4500/5000 M only).
To start/stop the air pump.
The air pump is off.
The air pump is on.
To unfreeze the screen after a finished measurement.
The screen is frozen.
The screen is unfrozen.