17 Service Utilities
108
C76IB003EN-E
Calculations
You can view a list of the calculations of the currently available output quantities.
DCB - Air Pump
DMA Density - Air Pressure
DCB - Environment
DCB - CAN Connector
To print live raw data
1. Tap <Menu>, select
"Service > Live Raw Data"
and choose one of the
groups of live raw data quantities.
2. To perform a printout on paper or to a PDF file, tap <Print or Export> and
follow the instructions on the screen.
RHOg5 Value
Viscosity corrected density, valid if the
viscosity is greater than 500 mPa·s.
Status Value
Status information indicating the
temperature status of the cell.
Air Pump State
Status of the air pump (0 or 1).
Air Pressure
Currently measured air pressure.
Air Pressure Condition
Current status of the measurement.
Air Pressure State
Status of the air pressure measurement.
Air Pressure Substate
Detailed information about the status of
the air pressure measurement.
Heatsink temperature
Temperature of the heat sink.
Room temperature
Temperature measured within the
instrument.
NTC temperature
Temperature measured within the
instrument.
CAN Connector Current
Current measured at the CAN interface.