12 Defining and Using Methods
C76IB003EN-E
65
4. Define the "Timeout".
5. Use the drop-down box "FillingCheck
TM
" to select one of the options:
•
always active
•
active during a measurement
•
not active
6. Tap <OK>.
12.2.2
Defining the Displayed Output Fields
For each method, you can select the number and content of output fields on the
display.
TIP
We recommend having always the output field "Density Condition" in addition to
the measurement values on the screen.
To define the number of output fields
When switching to another display layout, the contents of the output fields of the
previous display layout are transferred to the new layout.
1. Tap <Menu> and select
"Methods > Method Settings > "Method Name" >
Display Layout"
.
2. Highlight a list item and tap <OK>.
To define the content of output fields
You can directly change the content of output fields in the main screen. The
changes are automatically saved in the current method.
1. Tap on the output field.
2. Select the "Group", "Quantity", "Unit" and number of "Digits" and tap <OK>.
TIP
Before you use the output quantity "Canadian Excise Alcohol Table" of the group
"Ethanol Tables" (see Appendix F), it is necessary to perform a special
adjustment with apparent density values of air and water (see Chapter 11.3.5.2)
and to link that adjustment to the Canadian excise alcohol table (see Chapter
12.2.3
Defining the Result Output
You can define the list of output quantities for printouts and data exports. You can
change the settings any time and repeat the printout of saved results, if required.
If no output quantities are selected, the data browser settings are used instead
(see Chapter 15.2).